High-quality Leak Testing
The P-Test Plus 15 and the P-Test Plus 60 are used to perform Air Pressure Decay and Leak Rate testing on medical devices.
These Tecnoideal testing devices do well on a range of elastic or rigid medical devices such as blood lines, infusion and transfusion sets, bags, filters, and oxygenators.
Quality Assurance and R&D departments favor these testing tools for validation and qualification purposes, while the Production departments use them for routine testing of finished products.
The P-Test Plus 15 and P-Test Plus 60 can store up to 100 test programs, all protected by an alphanumeric 16-digit code.
These machines test pressure range from 30 to 6000 hPa/mBar. It records metrics for each program, including the number of tests executed, the pass/fail rate, and the tests’ date and time.
The P-Test Plus 15 and PT-Plus 60’s operative interface and technical features are easy to use. It possesses a wide LCD, backlit graphic display with contrast adjustment, and a user-friendly keyboard.
The P-Test Plus series arrives pre-calibrated and comes with a Calibration Certificate. The date of the last calibration performed is stored in the device’s memory and automatically displayed each time the machine starts. All devices include on-screen directions to help with an on-site calibration. Also, you can leverage Tecnoideal America Calibration Services.
Medical device manufacturers can pair the P-Test Plus 15 and 60 with optional value-add accessories, such as the 6-Way Sequencer, Bag Holder, Data Recorder, Remote Control, and Foot Switch.
Safety Systems
Operation & Quality Systems
P-Test Plus 15 | P-Test Plus 60 |
Range of Test Pressure 30 – 1500 hPa/mBar 3.00 – 150.00 kPa 23 – 1125.0 mmHg 310 – 15300 mmH2O 0.44 – 21.76 PSI | Range of Test Pressure 120 – 6000 hPa/mBar 12.00 – 600.00 kPa 90 – 4500 mmHg 1220 – 61180 mmH2O 1.74 – 87.05 PSI |
Decay Resolution 0.05 hPa | Decay Resolution 0.1 hPa |
Test accuracy | 0.25% of full scale |
Power Supply | 110 – 240V, 50-60Hz, 40 VA |
Electrical Protection | Electromagnetic interface filter and 2 1AT fuses |
Operating Temperature | From 10 to 40° |
Operating room humidity | Max. 80% |
Fuse | 2 1AT Fuses |
Inputs/Outputs | 1 RS232/485 opto-isolated1 RS232 for PC & Printer Connection1 interface for Remote Start or Foot Switch Connection |
Serial Ports | USB |
PLC Interface | For Program Selection, test start and test resultsPNP, isolated, 10-30Vdc; 8-pin connector (start/stop/passed/failed) |
Configurable Units of Measure | hPa(0mBar), kPa, mmHg, mmH2O, PSI |
Memory | Up to 100 test programs, with date, time, and alphanumeric 16-digit code |
Accessory Options | Foot switch, Remote Control, Bag Holder, 6-Way Sequencer, Data Recorder |